Start-Up and Use
Note: As the battery is used and recharged over a period of
time, the amount of time between the onset of the low
battery alarm and the instrument shut-off may become
shorter.
It is recommended that the internal battery be replaced by
qualified service personnel every 24 months.
Caution: If the NPB-290 is to be stored for a period of 2
months or longer, remove the battery from the monitor prior
to storage. Recharge the battery when it has not been
charged for 2 or more months.
DISPOSAL OF DEVICE COMPONENTS
Caution: Follow local governing ordinances and recycling
instructions regarding disposal or recycling of device
components, including batteries.
PERFORMANCE CONSIDERATIONS
Impact of Patient Conditions on Monitor Readings
Certain patient conditions can affect the measurements of the
NPB-290 and cause the loss of the pulse signal.
WARNING: Pulse oximetry readings and pulse signals can
be affected by certain ambient environmental conditions,
sensor application errors, and certain patient conditions.
Inaccurate measurements can be caused by:
• excessive patient movement
• venous pulsations
• intravascular dyes, such as indocyanine green or methylene
blue
• significant levels of dysfunctional hemoglobins
• defibrillation
Ambient environmental conditions and sensor application errors,
which can affect pulse oximetry readings, are discussed in the