Appendix B: Data Port Protocol
Table 3: Status Codes
Code Meaning
AO Alarm Off
AS Alarm Silence
LB Low Battery
LP Loss of Pulse
MO Patient Motion
PH Pulse Rate High
Limit Alarm
PL Pulse Rate Low
Limit Alarm
PS Pulse Search
SH Sat High Limit
Alarm
SL Sat Low Limit
Alarm
SD Sensor Disconnect
Note: A sensor disconnect will also cause three dashes (- - - ) to
be displayed in the patient data section of the printout.
TREND DATA PRINTOUT
The format of data displayed when a trend printout is requested is
similar to that of the real time data. The only differences are that
“TREND” is displayed in the top row instead of the “CRC:
XXXX” software verification number and there is no “Status”
column.
Readings are displayed in 2-second intervals. The values on each
row are an average for the 2-second period.
At the end of the printout an “Output Complete” line indicates
that the transmission was successful. If the “Output Complete”
line is not present, a corruption of the data may have been
detected and the data should be ignored.