OPERATOR’S MANUAL
NPB-295 Pulse Oximeter
Caution: Federal law (U.S.) restricts this device to sale by or on the order of a physician.
To contact Mallinckrodt’s representative: In the United States, call 1.800.635.5267
or 314.654.2000; outside of the United States, call your local Mallinckrodt representative.
0123
2002 Mallinckrodt Inc. All rights reserved. 061089C-1202
Zobrazit stránku 0
1 2 3 4 5 6 ... 86 87

Shrnutí obsahu

Strany 1 - OPERATOR’S MANUAL

OPERATOR’S MANUAL NPB-295 Pulse Oximeter Caution: Federal law (U.S.) restricts this device to sale by or on the order of a physician. To cont

Strany 2

Introduction Specific information about ambient environmental conditions, sensor application, and patient conditions is contained throughout this manu

Strany 3 - CONTENTS

CONTROLS, INDICATORS, AND SYMBOLS Displays, Controls, Indicators, and Connectors NPB-295 Symbols Description of Controls Description of Displays and

Strany 4

Controls, Indicators, and Symbols 035554A-0497NPB-2950123IPX1RNRTL/CTMT 0.50A 250V2XNELLCOR PURITAN BENNETT EUROPE BV,'s-HERTOGENBOSCH, THE NETHE

Strany 5

Controls, Indicators, and Symbols DESCRIPTION OF CONTROLS Function Buttons The Power On/Off Button. Used to turn the NPB-295 monitor on or off. Th

Strany 6 - FIGURES

Controls, Indicators, and Symbols The Pulse Amplitude Indicator (blip bar). Indicates pulse beat and shows the relative pulse amplitude. As the det

Strany 7 - SAFETY INFORMATION

Controls, Indicators, and Symbols Valid Button Press A short, medium-pitched tone indicating that an appropriate button has been pressed Invalid Butt

Strany 9

SETUP Unpacking and Inspection Performance Verification List of Components Monitor Setup Language Selection Tilt-Stand Supports UNPACKING AND INSPEC

Strany 10 - Introduction

Setup Optional Accessories A wall mount plate, as described in the Sensors and Accessories section is available from Mallinckrodt for use with this mo

Strany 11 - 101112141516 13

Setup Connecting The NPB-295 The NPB-295 operates on AC power when the hospital-grade power cord is connected to both the monitor and an AC power sour

Strany 12 - T 0.50A 250V

Nellcor Puritan Bennett Inc. 4280 Hacienda Drive Pleasanton, CA 94588 USA Telephone Toll Free 1.800.NELLCOR Mallinckrodt Europe BV Hambakenwetering

Strany 13 - DESCRIPTION OF CONTROLS

Setup 3. Plug the male connector end of the power cord into a properly grounded AC outlet. 4. Verify that the AC Power Indicator is lit. If it is n

Strany 14

Setup select the desired language. Press the EXIT button to return to the main menu. Service personnel may set the appropriate language as the power-

Strany 16 - (Blank Page)

SENSORS AND ACCESSORIES Selecting a Sensor Biocompatibility Testing Performance Considerations Wall Mount Plate SELECTING A SENSOR WARNING: Before

Strany 17

Sensors and Accessories Table 1: Nellcor Sensors Sensor Model Patient SizeOxisensor® II oxygen transducer (sterile, single-use only) N-25/N-25LF

Strany 18 - Optional Accessories

Sensors and Accessories PERFORMANCE CONSIDERATIONS WARNING: Pulse oximetry readings and pulse signal can be affected by certain ambient environmental

Strany 19 - Connecting The NPB-295

Sensors and Accessories WARNING: Tissue damage can be caused by incorrect application or duration of use of an SpO2 sensor. Inspect the sensor site

Strany 20

Sensors and Accessories The plate attaches to the bottom of the NPB-295 monitor as shown in Figure 4. If the tilt-stand supports are attached, remove

Strany 21

(Blank Page)

Strany 22

START-UP AND USE Basic Operation Alarms Adjustable Settings Menu Battery Operation Disposal of Device Components BASIC OPERATION WARNING: The NPB-2

Strany 23 - SENSORS AND ACCESSORIES

CONTENTS Figures Tables Contents... iii Figures...

Strany 24 - Sensors and Accessories

Start-Up and Use Ensure that the supply voltage selector switch on the rear panel is set to the proper voltage. Power-On Self-Test (POST) WARNING: En

Strany 25

Start-Up and Use 6. If a sensor is connected to the monitor and the patient, the Pulse Search Indicator lights and the NPB-295 displays zeroes in the

Strany 26

Start-Up and Use Monitoring Mode In Monitoring Mode - Pleth View (Figure 5), the NPB-295 displays SpO2 readings, pulse rate readings, and a pleth wave

Strany 27

Start-Up and Use Pulse Search If the acquired pulse is lost during monitoring, the NPB-295 enters Pulse Search. During Pulse Search, the monitor atte

Strany 28

Start-Up and Use When all of the following conditions are present for 15 minutes, the NPB-295 automatically shuts down: • Running on battery power •

Strany 29 - START-UP AND USE

Start-Up and Use • 50 seconds after loss of pulse is detected in the presence of continuous motion When operating on DC power, during a low battery

Strany 30 - Power-On Self-Test (POST)

Start-Up and Use SPORADIC OR LOST PULSE WITH CONTINUOUS MOTIONSPORADIC OR LOST PULSE WITHOUT CONTINUOUS MOTIONPulse Search Indicator lights continuous

Strany 31 - Contrast

Start-Up and Use During that data evaluation period: • If the patient is not moving (for example, because of cardiac arrest, shock, paralysis, anesth

Strany 32 - Monitoring Mode

Start-Up and Use ADJUSTABLE SETTINGS The following adjustments can be made using the Adjust Up/Down and Alarm Silence Buttons. • Pulse beep volume •

Strany 33

Start-Up and Use (If the button is pressed for more than 3 seconds or the duration is changed, the Alarm Silence Button must be pressed again to start

Strany 34 - Description of Alarms

Contents Management Of Loss-Of-Pulse Alarm... 29 Adjustable Settings ...

Strany 35

Start-Up and Use The factory default is that the capability of setting the alarm silence duration to OFF is enabled. Alarm Silence Reminder The alarm

Strany 36

Start-Up and Use menus, which will timeout in 5 minutes and the analog port calibration menu, which times out in 2 minutes. LIMITS TREND SETUP LIGHTS

Strany 37

Start-Up and Use LIMITS WARNING: Each time the monitor is used, check alarm limits to ensure that they are appropriate for the patient being monitore

Strany 38 - Alarm Silence Duration

Start-Up and Use EXITSELECT%SP02BPM96.79ALARM LIMITS%SPO2UPPERLOWERBPM10080.17040 Figure 9: Alarm Limits Selection TREND The NPB-295 can graphically

Strany 39 - Disabling Audible Alarms

Start-Up and Use EXITVIEW ZOOM NEXT12HR TREND%SPO225JUL98 14 : 35 : 0510090807095 : 98Figure 10: SpO2 Trend Periods of time when no measurements

Strany 40 - Menu Structure

Start-Up and Use For example, suppose the time represented by the right-hand edge of the screen in Figure 10 is 14:54:05. Scrolling one time period t

Strany 41

Start-Up and Use For example, in Figure 11, the histogram is for the last 12 hours. During those 12 hours, 68% of the %SpO2 measurements were from 96

Strany 42

Start-Up and Use Next The NEXT softkey gives you access to the DELETE and PRINT softkeys. Delete Pressing the DELETE softkey presents two options: YES

Strany 43 - EXITSELECT

Start-Up and Use • data port analog voltage. Press the SETUP softkey once to display VIEW and CLOCK. Then press NEXT to display COMM and LANG. Pres

Strany 44

Start-Up and Use have an ESC/P compatible graphical serial printer is attached to the NPB-295 serial port. Select OXINET if you are connecting to an

Strany 45 - Histogram

Contents Physical Characteristics ... 60 Weight ...

Strany 46 - 20-second trends table

Start-Up and Use LIGHT The LIGHT softkey allows you to turn the backlight on or off. Default Settings The NPB-295 is shipped with factory default sett

Strany 47

Start-Up and Use Nurse Call Feature WARNING: The nurse call feature should not be used as the primary source of alarm notification. The audible and

Strany 48

Start-Up and Use Since the monitor cannot operate with a fully discharged battery, before attempting to turn on an NPB-295 whose battery charge has be

Strany 49

Start-Up and Use Note: As the battery is used and recharged over a period of time, the amount of time between the onset of the low battery alarm and

Strany 51 - Nurse Call Feature

TROUBLESHOOTING AND MAINTENANCE Troubleshooting EMI Interference Obtaining Technical Assistance Returning the NPB-295 Maintenance TROUBLESHOOTING WA

Strany 52 - Low Battery Indicator

Troubleshooting and Maintenance Table 3: Error Codes and Messages Error Code Error Message Action 4 LOW BATTERY The battery is discharged to a critic

Strany 53

Troubleshooting and Maintenance DISALLOWED ON BATTERY - An attempt to print or download data port information while operating on battery power has bee

Strany 54

Troubleshooting and Maintenance 2. One or more display elements or indicators do not light during the power-on self-test. • Do not use the NPB-295; c

Strany 55

Troubleshooting and Maintenance • Excessive environmental motion or electromagnetic interference may be preventing the NPB-295 from tracking the puls

Strany 56 - Other Messages

Contents FIGURES Figure 1: NPB-295 Front Panel Display (Pleth View)... 5 Figure 2: NPB-295 Rear Panel...

Strany 57 - Suggested Corrective Actions

Troubleshooting and Maintenance units, cellular phones, mobile two-way radios, electrical appliances), it is possible that high levels of such interfe

Strany 58

Troubleshooting and Maintenance The software version appears in the monitor display each time the monitor successfully completes the power-on self-tes

Strany 59

Troubleshooting and Maintenance Cleaning WARNING: Do not spray, pour, or spill any liquid on the NPB-295, its accessories, connectors, switches, or o

Strany 60

SPECIFICATIONS Performance Electrical Environmental Conditions Physical Characteristics PERFORMANCE Measurement Range SpO2 0–100% Pulse Rate 20–250

Strany 61 - Performance Verification

Specifications Pulse Rate 20–250 bpm ±3 bpm Note: Pulse Rate accuracy is expressed as ±3 bpm across the display range. This variation equals ± one

Strany 62 - Cleaning

Specifications Battery Capacity A minimum of 8 hours with a new, fully charged battery under the following conditions: no alarms, no analog or serial

Strany 63 - SPECIFICATIONS

Specifications Operation Temperature +5°C to +40°C (+41°F to +104°F) Altitude/Barometric Pressure -390 m to 3,658 m (-1280 ft. to 12,000 ft.) +1060 hP

Strany 64

Specifications Mode of Operation Continuous 61

Strany 65

(Blank Page)

Strany 66

APPENDIX A: QUICK GUIDE TO OPERATION Introduction Settings Adjustments INTRODUCTION This Quick Guide to Operation is intended for use by experience

Strany 67 - Continuous

SAFETY INFORMATION General Safety Information GENERAL SAFETY INFORMATION This section contains important safety information related to general use o

Strany 68

Appendix A: Quick Guide to Operation Table 4: Settings Adjustments To Adjust Action Button Alarm Limits Press Press (to select parameter) Press

Strany 69 - INTRODUCTION

Appendix A: Quick Guide to Operation Table 4: Settings Adjustments To Adjust Action Button Time and Date Settings Press Press Press Press (to s

Strany 70

Appendix A: Quick Guide to Operation Table 4: Settings Adjustments To Adjust Action Button Language Press Press Press Press Press SETUP NEX

Strany 71

APPENDIX B: PRINCIPLES OF OPERATION Oximetry Overview OXIMETRY OVERVIEW Pulse oximetry is based on two principles: that oxyhemoglobin and deoxyhemo

Strany 72

Appendix B: Principles of Operation During monitoring, the instrument’s software reads this resistor and selects coefficients that are appropriate fo

Strany 73 - Automatic Calibration

Appendix B: Principles of Operation 010050 Saturation (%)pHTemperaturePCO22,3-DPGPO2 (mmHg)10050pHTemperaturePCO22,3-DPGFetal Hb Figure 13: Oxyhemogl

Strany 74

(Blank Page)

Strany 75 - Saturation (%)

APPENDIX C: DATA PORT PROTOCOL Overview Connecting to the Data Port Baud Rate Real-Time Display Format Trend Data Printout Nurse Call Analog Outputs

Strany 76

Appendix C: Data Port Protocol The cable used must have a braided shield providing 100% coverage, such as a Belden cable (Belden part number 9609) or

Strany 77 - . The cable should

Appendix C: Data Port Protocol 9 10111213141512345678 Figure 14: Data Port Pin Layout Pins 2, 3, and 5 provide data in RS-232 format. Pins 1, 4, 9, a

Strany 78 - . Do not create sharp bends

Safety Information Standard 601-1 for electromedical equipment. All combinations of equipment must be in compliance with IEC Standard 601-1-1 systems

Strany 79 - 12345678

Appendix C: Data Port Protocol 74 Data in the highlighted box above represents the model number of the monitor, in this case the NPB-295. A column hea

Strany 80 - Data Source

Appendix C: Data Port Protocol Software Revision Level The next data field tells the user the software level, (Version 2.0.0.0) and a software verifi

Strany 81

Appendix C: Data Port Protocol Patient Data Patient data is highlighted in the display above. Parameter values are displayed directly beneath the he

Strany 82 - Operating Status

Appendix C: Data Port Protocol SH Sat High Limit Alarm SL Sat Low Limit Alarm SD Sensor DisconnectNote: A sensor disconnect will also cause three

Strany 83 - . A trend

Appendix C: Data Port Protocol 2HR TREND 82: 83%SPO210070809010APR98 10:00:00 Figure 17: Trend Data Printout (GRAPH MODE) Trend Data Operation Whenev

Strany 84 - 10APR98 10:00:00

Appendix C: Data Port Protocol If battery power is disconnected or depleted, trend data and user settings will be lost. All data is stored with error

Strany 85

Appendix C: Data Port Protocol Table 7: Voltage between pins 10 and 11 Alarm State Nurse Call Polarity Setting Voltage from pins 10 to 11 No current

Strany 86

Appendix C: Data Port Protocol The analog function can be accessed from the main menu by pressing SETUP, NEXT, NEXT, ANALOG. Selecting “0 VOLT” or “1

Strany 87 - . Selecting “0 VOLT”

INTRODUCTION Intended Use General Operating Principles and Conditions INTENDED USE The purpose and function of the Nellcor NPB-295 pulse oximeter is