Start-Up and Use
have an ESC/P compatible graphical serial printer is attached to
the NPB-295 serial port. Select OXINET if you are connecting to
an Oxinet® II central station network or Score software. Do
not use the CLINICAL setting unless instructed to do so by
Mallinckrodt customer service personnel. Press Exit to return to
the SETUP submenu.
Note: When using Score software use the latest version.
Contact Mallinckrodt’s Technical Services
Department or your local Mallinckrodt representative
to determine the latest version of Score software.
LANG
The LANG softkey allows you to select the language displayed
on the screen.
Press the NEXT softkey to access the LANG softkey. After
pressing the LANG softkey, use the Adjust Up/Down Buttons to
select Dutch, English, French, German, Italian, Portuguese, or
Spanish.
If the language is changed and EXIT is pressed, (or a 10-second
timeout occurs), the monitor begins displaying data in the
selected language.
NCALL
The NCALL softkey allows you to set the alarm voltage at a
normally high (NORM +) or normally low (NORM -). Refer to
the “Nurse Call” heading of Appendix C: Data Port Protocol for
a more thorough explanation of these settings. The nurse call
feature is also discussed later in this section.
ANALOG
The ANALOG softkey allows you to calibrate the NPB-295 data
port voltages with instruments such as a chart recorder. Refer to
the “Analog Outputs” heading of Appendix C: Data Port
Protocol for a more thorough explanation of these settings.
43