Start-Up and Use
The factory default is that the capability of setting the alarm
silence duration to OFF is enabled.
Alarm Silence Reminder
The alarm silence reminder (three beeps) will sound at least every
3 minutes while the alarm silence duration is set to OFF and the
Alarm Silence Indicator is flashing.
The alarm silence reminder may also be disabled by service
personnel using the procedures described in the service manual.
The factory default is that the reminder is enabled.
MENU
Menu Structure
The four softkeys on the front panel are used to view or adjust the
following NPB-295 settings or functions:
• %SpO
2 and pulse rate alarm limits
• Pleth or magnified view
• Time and date settings
• Data port baud rate settings
• Trend data viewed (%SpO
2, pulse, or both)
• “Zoom” factor of trend data
• Graph of trend data (histogram)
• Delete all trend data
• Print trends
• Language displayed on screen or data port
• Nurse call settings
• Analog output settings
Menu items are selected by pressing and releasing the
corresponding softkey directly below the item. Refer to Figure 8
to access menu items.
Note: If, after accessing a submenu, no buttons are pressed for
10 seconds, the display will timeout and return to the
main menu. Exceptions to this are the clock and trend
34