Troubleshooting and Maintenance
The battery should be replaced at least every 24 months. Refer to
the NPB-290 service manual for the battery changing procedure.
If service is necessary, contact qualified service personnel or your
local Mallinckrodt representative.
Periodic Safety Checks
It is recommended that the following checks be performed every
24 months:
Inspect the equipment for mechanical and functional
damage.
Inspect the safety relevant labels for legibility.
Performance Verification
If the monitor has been visibly damaged or subjected to
mechanical shock (for example, when dropped), qualified service
personnel should perform the procedure in the Performance
Verification section of the NPB-290 service manual.
Cleaning
WARNING: Do not spray, pour, or spill any liquid on the
NPB-290, its accessories, connectors, switches, or openings in
the chassis.
The NPB-290 may be surface-cleaned by using a soft cloth
dampened with either a commercial, nonabrasive cleaner or a
solution of 70% alcohol in water, and lightly wiping the surfaces
of the monitor.
The surface of the monitor may be disinfected using a soft cloth
saturated with one of the following solutions:
10% chlorine bleach in water
gluteraldehyde (cidex or equivalent)
Before attempting to clean an SpO
2 sensor, read the directions for
use enclosed with the sensor. Each sensor model has cleaning
instructions specific to that sensor.