Data Port Protocol
92
Data Source
N-395 VERSION 2.0.0.0 CRC: XXXX SpO2 Limit: 70-100% PR Limit: 60-160BPM
TIME %SpO2 BPM PA Status
Data in the highlighted box above represents the model number
of the monitor, in this case the N-395.
Software Revision Level
N-395 VERSION 2.0.0.0 CRC: XXXX SpO2 Limit: 70-100% PR Limit: 60-160BPM
TIME %SpO2 BPM PA Status
The next data field tells the user the software level, (Version
2.0.0.0) and a software verification number (CRC: XXXX).
Neither of these numbers should change during normal operation.
The numbers may change if the monitor is serviced and receives
a software upgrade.
Alarm Limits
N-395 VERSION 2.0.0.0 CRC: XXXX SpO2 Limit: 70-100% PR Limit: 60-160BPM
TIME %SpO2 BPM PA Status
The last data field in the top line indicates the high and the low
alarm limits for %SpO
2 and for the pulse rate (PR). In the
example above the low alarm limit for SpO
2 is 70% and the high
alarm limit is 100%. Pulse Rate alarm limits are 60 and 160
bpm.
Column Headings
N-395 VERSION 2.0.0.0 CRC: XXXX SpO2 Limit: 70-100% PR Limit: 60-160BPM
TIME %SpO2 BPM PA Status
Actual column headings are in the second row of the column
heading line. Patient data presented in the chart, from left to
right, is the:
time the patient data was obtained
current %SpO
2 value
current Pulse Rate
current Pulse Amplitude
operating status of the N-395.