Data Port Protocol
96
alarm determine the voltage between these pins. The nurse call
polarity is set by using the procedures in the Start-up and Use
section. To access the nurse call menu from the main menu,
press softkeys SETUP, NEXT, NEXT and NCALL.
When the nurse call polarity setting is normally high (NORM +)
and there is no alarm condition, the voltage between pins 11 and
10 will be +5 to +12 volts DC. Whenever the monitor is in an
alarm condition, the output between pins 11 and 10 will be -5 to
-12 volts DC. When the setting is normally low (NORM -), the
readings are opposite, as indicated in Table 12.
These voltages are present only when the monitor is operating on
AC power.
If the audible alarm has been turned off, or silenced, the nurse
call alarm is also silenced.
Table 12: Voltage Between Pins 10 and 11
Alarm State Nurse Call
Polarity Setting
Voltage from pins
10 to 11
No current alarms Normally high +5 to +12 VDC
Alarm condition Normally high -5 to -12 VDC
No current alarms Normally low -5 to -12 VDC
Alarm condition Normally low +5 to +12 VDC
Pins 7 and 15 provide a solid state relay that closes when an
alarm is sounding on the monitor. Pins 8 and 15 provide a solid
state relay that opens when an alarm is sounding. Pin 15 is a
common lead for both relays. The solid state relay operates
whether the monitor is operating on AC power or battery.