Start-Up and Use
62
Note: If the AC voltage selector switch on the rear panel does
not match the AC voltage source, the monitor may run
on battery power, even though it is plugged in, which
will eventually result in a low priority alarm and a
lighted low battery indicator. Ensure that the switch
setting matches the AC voltage.
If the monitor is not connected to AC power within
approximately 15 minutes, it will shut down.
Note: As the battery is used and recharged over a period of
time, the amount of time between the onset of the low
battery alarm and the instrument shut-off may become
shorter.
If the backlight is turned off during a low battery condition, it
cannot be turned back on.
It is recommended that qualified service personnel replace the
internal battery every 24 months.
Caution: If the N-395 is to be stored for a period of 3
months or longer, notify service personnel to remove the
battery from the monitor prior to storage. Recharge the
battery when it has not been charged for 2 or more months.
DISPOSAL OF DEVICE COMPONENTS
Caution: Follow local governing ordinances and recycling
instructions regarding disposal or recycling of device
components, including batteries.
PERFORMANCE CONSIDERATIONS
Impact of Patient Conditions on Monitor Readings
Certain patient conditions can affect the measurements of the
N-395 and cause the loss of the pulse signal.
WARNING: Pulse oximetry readings and pulse signals can
be affected by certain ambient environmental conditions,
sensor application errors, and certain patient conditions.