Data Port Protocol
94
Table 11: Status Codes
Code Meaning
AO
Alarm Off
AS
Alarm Silence
LB
Low Battery
LM
Loss of Pulse w/ Motion
LP
Loss of Pulse
MO
Patient MOtion
PH
Pulse Rate High Limit Alarm
PL
Pulse Rate Low Limit Alarm
PS
Pulse Search
SH
Sat High Limit Alarm
SL
Sat Low Limit Alarm
SD
Sensor Disconnect
Note: A sensor disconnect will also cause three dashes (- - -) to
be displayed in the patient data section of the display or
printout.
TREND DATA PRINTOUT (ASCII MODE)
The format of data displayed when a trend printout (Figure 24) is
requested is similar to that of the real-time data. The only
differences are that “TREND” is displayed in the top row instead
of the “CRC: XXXX” software verification number and there is
no “Status” column.
Readings are displayed in 4-second intervals. The values on each
row are an average for the 4-second period.
At the end of the printout an “Output Complete” line indicates
that the transmission was successful. If the “Output Complete”
line is not present, a corruption of the data may have been
detected and the data should be ignored.